GE Healthcare
Revision 1
Job Card D/R A232 - Lift Screw Assembly
Gantry and place them in close proximity to the DC motor.
15. Release the four screws (14) that connect the lower part (Bearing Block) of the Lift Screw Assembly
to the Gantry frame (5 mm allen wrench). While releasing these screws, carefully hold the base (15)
of the Lift Screw Assembly so that it does not fall.
Chapter 9
Service Information and Procedures Class A 2385072-16-8EN
15
14
Page no. 1264
13
Senographe DS
JC-DR-A-232.fm