GE Senographe DS Manual page 1680

Hide thumbs Also See for Senographe DS:
Table of Contents

Advertisement

GE Healthcare
Revision 1
Job Card CAL A021 - Arm Rotation Calibration
6-2-1
Set the 0° position
1. You are asked to move the arm to the 0° position. Press and hold the appropriate rotation movement
control button. Stop the arm at the 0° position.
2. Check that the arm is in the correct position (0° ±0.5°) by using a spirit level or similar device to check
that the top of the digital detector is horizontal.
3. If necessary, use the movement control buttons to adjust the arm as required until it is in the 0° posi-
tion.
4. When the arm has been correctly positioned, press the collimator FOV size button (behind the tube
head) to signal completion of the 0° adjustment.
6-2-2
Set the 185° position
1. You are now asked to move the arm to the 185° position. Press and hold the appropriate rotation
movement control button to rotate the arm counter-clockwise until it stops.
2. Check that the arm is in the correct position (185° ±0.5°) by using a spirit level and a suitable angle
measuring device to check that the top of the digital detector (now pointing down) is at 5° from the
horizontal.
3. If necessary, use the movement control buttons to adjust the arm as required until it is in the 185°
position.
4. When the arm has been correctly positioned, press the collimator FOV size button (behind the tube
head) to signal completion of the 185° adjustment.
5. If the 185° position cannot be attained, or the message on the Gantry readout indicates an error,
repeat the procedures to set the 0° and 185° positions. If the problem persists, it will be necessary to
investigate for a hardware error.
6-2-3
Set the -165° position
1. You are now asked to move the arm to the -165° position. Press and hold the appropriate rotation
movement control button to rotate the arm clockwise until it stops.
2. Check that the arm is in the correct position (-165° ±0.5°) by using a spirit level and a suitable angle
measuring device to check that the top of the digital detector (now pointing down) is at 15° from the
horizontal.
3. If necessary, use the movement control buttons to adjust the arm as required until it is in the -165°
position.
4. When the arm has been correctly positioned, press the collimator FOV size button (behind the tube
head) to signal completion of the -165° adjustment.
5. If the -165° position cannot be attained, or the message on the Gantry readout indicates an error,
repeat the procedures to set the 0°, 185° and -165° positions. If the problem persists, it will be neces-
sary to investigate for a hardware error.
6-2-4
Completion
If the calibration has been successfully completed, save the results by clicking the Save button on the
AWS Screen.
If you wish to exit the calibration without saving the results (leaving the system un-calibrated or using
the previous settings), click the Cancel button on the AWS Screen.
Chapter 9
Service Information and Procedures Class A 2385072-16-8EN
Page no. 1680
Senographe DS
JC-CAL-A-021.fm

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Senographe DS and is the answer not in the manual?

Questions and answers

Dr.Robert Mahmut
March 26, 2025

G.E.Essential l have recently bought What is the full weight.of the mamo Full weight of the console and other İtem's also measurements of each item with the main machine please All this Shipping Company required

1 comments:
Mr. Anderson
March 26, 2025

The GE Senographe DS mammography machine is shipped in three crates with the following dimensions and weights:

- Crate 1 (Gantry, Generator Cabinet, and ramp):
- Dimensions: 2066 mm (81.34 in) depth × 848 mm (33.39 in) width × 2292 mm (90.24 in) height
- Mass: 641 kg (1410.2 lbs)

- Crate 2 (Control Station):
- Dimensions: 1110 mm (43.70 in) depth × 770 mm (30.31 in) width × 1850 mm (72.83 in) height
- Mass: 210.5 kg (464.07 lbs) for truck transport (without side covers and lid)
- Mass: 223.5 kg (492.73 lbs) for plane/boat transport (with side covers and lid)

- Crate 3 (Detector):
- Dimensions: 872 mm (34.33 in) depth × 2270 mm (89.37 in) width × 260 mm (10.24 in) height
- Mass: 104.4 kg (229.68 lbs)

### Total Mass:
- For truck transport: 641 + 210.5 + 104.4 = 955.9 kg (2104.95 lbs)
- For plane/boat transport: 641 + 223.5 + 104.4 = 968.9 kg (2133.61 lbs)

These values include the gantry, generator cabinet, control station, detector, and all related components.

This answer is automatically generated

Table of Contents