GE Healthcare
Revision 1
Job Card D/R A264 - Tilt Sensor
1
SUPPLIES
Zero Tilt Sensor and Cable W405
2
TOOLS
Standard Tool Box
3
REQUIRED EFFORT
Personnel: 1 Field Engineer
Time: 2 hours
4
SAFETY PRECAUTIONS
WARNING
The Gantry must be powered off.
5
PREREQUISITES
•
Rotate the arm to 90° for access to the Tilt Sensor (1).
•
Switch off the Gantry power.
•
Remove the tube cover (refer to
Remove/Reinstall Gantry Covers on page
JC-DR-A-264.fm
Service Information and Procedures Class A 2385072-16-8EN
Job Card PHY A044 -
523).
Page no. 1357
Job Card D/R A264 - Tilt Sensor
1
Senographe DS
Chapter 9
Chapter 9
Need help?
Do you have a question about the Senographe DS and is the answer not in the manual?