4. Ensure that the JP3 and JP4 (2) jumpers and JP1 and JP2 (3) above and below the XJ8 connector
on the Generator Command board are set the same as the defective board that you have just
replaced.These jumpers determine the room lamps and door configuration, which can vary one each
customer site.
5. If the customer wants a different lamp and door configuration to before, set jumpers JP1 and JP2 as
required according to the table below. Place unused jumpers in the parking positions JP3 and JP4.
Note:
If there is no door switch, jumpers JP1 and JP2 must both be present (case 4 in the table).
JC-DR-A-406.fm
Description of cable
(usually identified by code on a white label)
(red line on right of ribbon connector)
(ribbon cable to kVmA board, red line on right of ribbon connector)
(red line on right of ribbon connector)
(red line on top of ribbon connector)
The X-ray Console cable
400 PL2 XJ2, 45440116, 400 PL1 X10
(red line on top of ribbon connector)
(red line on left of ribbon connector)
Service Information and Procedures Class A 2385072-16-8EN
Job Card D/R A406 - Generator Command Board 400-PL1
400 PL1 XJ1
XJ2
400 PL1 XJ3
400 PL1 XJ4
C801418 XP1 XP2
400 PL1 XJ6
400 PL1 XJ7
45440115
Empty
400 PL1 XJ13
2
Page no. 1573
Senographe DS
Cable color
Gray
Black
Black
5 Black
1 White
Gray
Gray
Gray
White
Black
Gray
N/A
Gray
3