Job Card D/R A265 - Tilt Motor And Drive Assembly - GE Senographe DS Manual

Hide thumbs Also See for Senographe DS:
Table of Contents

Advertisement

GE Healthcare
Revision 1

Job Card D/R A265 - Tilt Motor and Drive Assembly

1
SUPPLIES
Entire Tilt Motor and Tilt Drive Assembly with Bearing
Or separately, as required:
Motor-Tilt + W404
Tilt Drive Assembly
Bearing-Tilt Drive
2
TOOLS
Standard Tool Box
3
REQUIRED EFFORT
Personnel: 1 Field Engineer
Time: 2 hours
4
SAFETY PRECAUTIONS
WARNING
The Gantry must be powered off.
5
PREREQUISITES
Rotate the arm to 90° for access to the Tilt Motor (1).
Switch off the Gantry power.
Remove the tube cover (refer to
Remove/Reinstall Gantry Covers on page
JC-DR-A-265.fm
(2345929)
(2392469)
(99033678)
Job Card PHY A044 -
Service Information and Procedures Class A 2385072-16-8EN
Job Card D/R A265 - Tilt Motor and Drive Assembly
(2345905)
523).