GE Healthcare
Revision 1
Job Card D/R A244 - Compression Motor with Encoder and Compression
Belt
1
SUPPLIES
•
1
Compression Motor
•
1
Compression Belt
2
TOOLS
•
Standard Tool Box
•
Hub extractor
•
Mammography compression scale with digital display
3
REQUIRED EFFORT
Personnel: 1 Field Engineer
Time: 45 minutes
4
SAFETY PRECAUTIONS
WARNING
The Gantry must be powered off.
5
PREREQUISITES
•
Lower the arm until you have good access to the Com-
pression Motor (1).
•
Switch off the Gantry power.
•
Remove the arm covers (refer to
Remove/Reinstall Gantry Covers on page
JC-DR-A-244.fm
Service Information and Procedures Class A 2385072-16-8EN
Job Card D/R A244 - Compression Motor with Encoder and Compression Belt
Job Card PHY A044 -
523).
Page no. 1297