Job Card D/R A313 - Trackball
6. Disconnect the trackball cable from the USB extension cable leading to the ADS computer.
7. Cut the cable ties that secure the monitor cables and trackball cable to the Control Station framework
(3) and harness (4).
4
3
8. Carefully pull the harness (5) from the upper and lower parts of the rotative arm.
7
5
9. Extract the trackball cable (6) from the hole in the harness, and from the opening of the monitor sup-
Chapter 9