GE Senographe DS Manual page 1504

Hide thumbs Also See for Senographe DS:
Table of Contents

Advertisement

GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Job Card D/R A321 - IDC Computer
6
PROCEDURE
6-1
Summary
The following list summarizes the steps described in this document.
These steps must be followed when installing a replacement IDC Computer:
1. Disassembly (removal) of the existing IDC Computer.
2. Reassembly (installation) of the replacement IDC Computer.
3. Reload IDC software from the IDC Application Software CD-ROM used for the current installation.
4. Restore the detector calibration files by applying a "Restore IDC", see
Job Card ELE A052 - Back-up/
Restore Procedure on page
771.
5. Select Service desktop/Utilities and perform the Detector Replacement procedure.
6. Reset the IDC Computer by powering it OFF/ON.
Chapter 9
Page no. 1504
JC-DR-A-321_DS.fm