Job Card D/R A401 - Generator Cpu Board 400Pl3 - GE Senographe DS Manual

Hide thumbs Also See for Senographe DS:
Table of Contents

Advertisement

GE Healthcare
Revision 1

Job Card D/R A401 - Generator CPU Board 400PL3

1
SUPPLIES
Generator CPU Board
2
TOOLS
None
3
REQUIRED EFFORT
Personnel: 1 Field Engineer
Time: 2 hours
4
SAFETY PRECAUTIONS
No specific safety precautions are applicable.
When handling circuit boards, take precautions against damage by ESD (electro-static discharge), by
using for example an anti-static mat and an ESD strap.
5
PREREQUISITES
CAUTION
Generator parameters are saved in the NV-RAM of the Generator CPU Board. If possible,
ensure that a backup of these parameters has been made before removing the board. If
not, it will be necessary to reconfigure the Generator after changing the board. To back-up
the NV-RAM of the Generator CPU Board, perform a full system back-up according to
Card ELE A052 - Back-up/Restore Procedure on page
For Senographe systems that use AWS V43.0 or later (on the SB150) or AWS 53.0 or later (on the U20),
or AWS 54.10 or later (on the Z400), consult the site System Configuration Form to determine the cur-
rent AOP configuration for all three modes (DOSE, STD, and CNT) on the Generator.
JC-DR-A-401.fm
Service Information and Procedures Class A 2385072-16-8EN
Job Card D/R A401 - Generator CPU Board 400PL3
Page no. 1543
771.
Senographe DS
Chapter 9
Job
Chapter 9

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Senographe DS and is the answer not in the manual?

Table of Contents

Save PDF