Job Card CAL A043 - Detector Gain Calibration
8
ERROR ACTIONS
If an error is found, an error message appears, with a request for further action. Possible error messages
are:
1. IQ check failed - Detector not ready. You are informed that the IQ check failed because the check for
detector lag failed, or because the detector temperature was outside limits. You are then requested
to try the test again after a 15 minute delay or after the detector temperature has stabilized.
2. IQ check failed - system configuration is not OK. You are asked to repeat the IQ check after checking
the system configuration (phantom position, removal of paddle, collimator in field of view).
9
COMPLETION
If you are handing the machine back to the user at this point, switch the Senographe off, then on again.
Chapter 9
Service Information and Procedures Class A 2385072-16-8EN
Page no. 1700