GE Senographe DS Manual page 1695

Hide thumbs Also See for Senographe DS:
Table of Contents

Advertisement

GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Job Card CAL A041 - Bad Pixel Calibration
tion specific to the detector.
For more information about Bad Pixel troubleshooting, refer to
Job Card TSG A010 - Troubleshooting
Bad Pixel Calibration Failure on page
1019.
6-4
Gain Calibration
After completing the Bad Pixel calibration, Gain Calibration must be performed for all techniques (per-
form
Job Card CAL A043 - Detector Gain Calibration on page
1697).
7
COMPLETION
If you are handing the machine back to the user at this point, switch the Senographe off, then on again.
Page no. 1695
Chapter 9
JC-CAL-A-041.fm

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents