Agilent Technologies 35670A Manual page 96

Dynamic signal analyzer gpib programming
Hide thumbs Also See for 35670A:
Table of Contents

Advertisement

ABORt
ABORt
Stops the current measurement in progress.
Command Syntax:
Example Statements:
Attribute Summary:
Description:
This command aborts any measurement in progress and resets the trigger system. ABOR forces the
trigger system to an idle state. ABORt forces the Measuring bit (bit 4) and the Averaging bit (bit 8) of the
Operation Status register to 0. To restart a measurement you must send the INIT:IMM command.
The program message ABOR;:INIT:IMM serves a special synchronization function. When you send this
message to restart a measurement, the analyzer's No Pending Operation (NPO) flag is set to 1 until the
measurement is complete. The two commands that test the state of this flag—*WAI and *OPC—allow
you to hold off subsequent actions until the measurement is complete. See chapter 3 of the GPIB
Programmer's Guide for more information about synchronization.
Note
When averaging is on ([SENSe:]AVERage[:STATe] ON) the NPO flag is not set to 1
until n measurements have been combined into one trace. You specify the value of n
with the[SENSe:] AVER:COUNt command.
4-2
ABORt
OUTPUT 711;"abor"
OUTPUT 711;"Abort"
Option:
not applicable
Synchronization Required:
Preset State:
not applicable
SCPI Compliance:
no
confirmed
command

Advertisement

Table of Contents
loading

Table of Contents