Agilent Technologies 35670A Manual page 325

Dynamic signal analyzer gpib programming
Hide thumbs Also See for 35670A:
Table of Contents

Advertisement

HCOPy:ITEM:LABel:TEXT
Specifies a label for plot and print output.abeling;plot/print
Command Syntax:
<STRING> ::= maximum of 119 ASCII characters
Example Statements:
Query Syntax:
Return Format:
Attribute Summary:
Description:
This command specifies a label for your plot or print output. The label appears at the top of the screen.
See figure 10-1 on page 10-25. The label is plotted (or printed) when HCOP:ITEM:LAB:STAT is ON
and the HCOP:IMM command is executed.
A line feed character (decimal value 10) in <STRING>, divides the label into two lines.
Specify a label with a space to have the area appear blank on the plot (the mini-state will not appear).
If you can use a command that does not comply with SCPI, see the HCOP:TITle command.
HCOPy:ITEM:LABel:TEXT '<STRING>'
59
characters per line
OUTPUT 711;":hcopy:item:lab:text 'Test 1 Results'
OUTPUT 711;"HCOP:ITEM:LABEL:TEXT "LINE1"&CHR$(10)&"LINE2"
HCOPy:ITEM:LABel:TEXT?
"<STRING>"
Option:
not applicable
Synchronization Required:
Preset State:
not applicable
SCPI Compliance:
no
approved
HCOPy
command/query
10-11

Advertisement

Table of Contents
loading

Table of Contents