STATus:QUEStionable:LIMit:ENABle
Sets and queries bits in the Limit Fail enable register.
Command Syntax:
<number> ::= a real number (NRf data)
<bound> ::= MAX|MIN
Example Statements:
Query Syntax:
Return Format:
Attribute Summary:
Description:
To set a single bit in the Limit Fail enable register to 1, send the bit's decimal weight with this command.
To set more than one bit to 1, send the sum of the decimal weights of all the bits. (The decimal weight of
n
a bit is 2
, where n is the bit number.)
All bits are initialized to 0 when the analyzer is turned on or when the STAT:PRES command is sent.
However, the current setting of bits is not modified when you send the *RST command.
All bits are initialized to 0 when the analyzer is turned on. However, the current setting of bits is not
modified when you send the *RST command.
See "Limit Fail Register Set" in chapter 1 for a definition of bits in the register set. See "Programming
the Status System" in the GPIB Programmer's Guide for information about the role of enable
registers in register sets.
STATus:QUEStionable:LIMit:ENABle <number>|<bound>
limits:
0:32767
OUTPUT 711;"stat:ques:lim:enab 3"
OUTPUT 711;"STATUS:QUES:LIMIT:ENABLE 15"
STATus:QUEStionable:LIMit:ENABle?
Integer
Option:
not applicable
Synchronization Required:
Preset State:
not affected by Preset
SCPI Compliance:
no
instrument-specific
STATus
command/query
20-17