[SENSe:]VOLTage[1|2|3|4][:DC]:RANGe:UNIT:USER:SFACtor
Specifies the transducer sensitivity for transducer units.
Command Syntax:
<number> ::= a real number (NRf data)
<unit> ::= [V/EU|EU/V]
<step> ::= UP|DOWN
<bound> ::= MAX|MIN
Example Statements:
Query Syntax:
Return Format:
Attribute Summary:
Description:
This command sets the sensitivity of a mechanical-to-electrical transducer as the ratio between the
electrical signal (output) and mechanical quantity (input). The ratio is Volts per transducer unit.
The transducer sensitivity and the transducer unit label are used only when
VOLT:RANG:UNIT:USER:STATe is ON.
If the channel specifier is not used, the command defaults to channel 1.
Refer to the documentation for your transducer for the appropriate sensitivity value.
[SENSe:]VOLTage[1|2|3|4][:DC]:RANGe:UNIT:USER:SFACtor
{<number>[<unit>]}|<step>|<bound>
limits:
-9.9e37:9.9e37
OUTPUT 711;"volt:rang:unit:user:sfac 0.1"
OUTPUT 711:"SENS:VOLTAGE3:DC:RANGE:UNIT:USER:SFACTOR
1.8e-06"
[SENSe:]VOLTage[1|2|3|4][:DC]:RANGe:UNIT:USER:SFACtor?
Real
Option:
not applicable
Synchronization Required:
Preset State:
not applicable
SCPI Compliance:
no
instrument-specific
[SENSe:]
command/query
18-83