Agilent Technologies 35670A Manual page 523

Dynamic signal analyzer gpib programming
Hide thumbs Also See for 35670A:
Table of Contents

Advertisement

[SENSe:]VOLTage[1|2|3|4][:DC]:RANGe:UNIT:XDCR:LABel
Specifies a transducer unit for the specified channel.
Command Syntax:
Example Statements:
Query Syntax:
Return Format:
Attribute Summary:
Description:
Select a unit label appropriate for your transducer. The transducer unit label appears only when
VOLT:RANG:UNIT:USER is ON.
This command works with the CALC:UNIT:MECH command. If transducer units are G, M/S2, M/S, M,
INCH/S2, INCH/S, INCH, or MIL; the analyzer automatically integrates or differentiates the input data
using the CALC:UNIT:MECH command. If the transducer unit is PA, the analyzer automatically
displays dBSPL as the Y-axis unit for dB magnitude coordinate transforms.
If your transducer measures in units other than pressure, acceleration,velocity, or displacement; you will
have to set this command to USER (VOLT:RANG:UNIT:XDCR:LAB USER) and define your own label
with the VOLT:RANG:UNIT:USER:LABel command.
If the channel specifier is not used, the command defaults to channel 1. The query returns the last-entered
transducer unit name for the specified channel.
[SENSe:]VOLTage[1|2|3|4][:DC]:RANGe:UNIT:XDCR:LABel
PA|G|M/S2|M/S|M|KG|N|DYN|INCH/S2|INCH/S|INCH|MIL|LB|USER
OUTPUT 711;"SENSE:VOLT2:RANG:UNIT:XDCR:LABEL PA"
OUTPUT 711;"volt:rang:unit:user:lab G"
[SENSe:]VOLTage[1|2|3|4][:DC]:RANGe:UNIT:XDCR:LABel?
PA|G|M/S2|M/S|M|KG|N|DYN|INCH/S2|INCH/S|INCH|LB|MIL|USER
Option:
not applicable
Synchronization Required:
Preset State:
USER
SCPI Compliance:
no
instrument-specific
[SENSe:]
command/query
18-85