This subsystem contains commands which provide access to the raw measurement data (data that has not
been transformed into the current display coordinates). The commands, TRAC:DATA and
TRAC:WAT:DATA, allow you to transfer measurement data between the analyzer and an external
controller.
Figure 23-1 shows you the position of TRAC:DATA in the data flow. It also illustrates the difference
between data available in the TRACe subsystem and the CALCulate subsystem.
After measurement data is collected, any specified math operations are performed. Data is then
transformed into the specified coordinate system and sent to the display. TRAC:DATA gives you access
to the raw measurement data after math operations have been performed. This data can be either complex
or real. CALC:DATA gives you access to the display data—after the coordinate transformation.
Note
Both TRAC:DATA and CALC:DATA allow you to transfer measurement data from the
analyzer. Only TRAC:DATA, however, allows you to transfer measurement data to the
analyzer.