CALCulate
CALCulate[1|2|3|4]:MARKer:MAXimum[:GLOBal]:TRACk
Turns the peak tracking function on or off.
Command Syntax:
Example Statements:
Query Syntax:
Return Format:
Attribute Summary:
Description:
When peak tracking is enabled, the analyzer automatically positions the main marker on the largest peak
of the specified trace each time the trace is updated.
The trace specifier, CALCulate[1|2|3|4], determines which trace you are selecting. Omit the specifier or
send 1 for trace A, 2 for trace B, 3 for trace C, or 4 for trace D. The trace does not need to be displayed
nor must it be the primary active trace.
To move the marker to the highest peak one time, use the CALC:MARK:MAX:GLOB command.
6-76
CALCulate[1|2|3|4]:MARKer:MAXimum[:GLOBal]:TRACk OFF|0|
ON|1
OUTPUT 711;"calculate2:mark:maximum:glob:trac OFF"
OUTPUT 711;"Calculate3:Mark:Maximum:Trac ON"
CALCulate[1|2|3|4]:MARKer:MAXimum[:GLOBal]:TRACk?
Integer
Option:
not applicable
Synchronization Required:
Preset State:
OFF (+0)
SCPI Compliance:
no
instrument-specific
command/query