Agilent Technologies 35670A Manual page 483

Dynamic signal analyzer gpib programming
Hide thumbs Also See for 35670A:
Table of Contents

Advertisement

[SENSe:]FREQuency:SPAN:LINK
Specifies the frequency parameter which remains constant if frequency span or record length is modified.
Command Syntax:
Example Statements:
Query Syntax:
Return Format:
Attribute Summary:
Description:
This command "anchors" or "fixes" the start frequency or the center frequency.
If FREQ:SPAN:LINK STAR is sent, the start frequency does not change when the frequency span or
record length changes.
Note
FREQ:SPAN:LINK is set to CENTer after a reset (*RST).
[SENSe:]FREQuency:SPAN:LINK STARt|CENTer
OUTPUT 711;":Freq:Span:Link CENTER"
OUTPUT 711;"SENS:FREQUENCY:SPAN:LINK CENTER"
[SENSe:]FREQuency:SPAN:LINK?
STAR|CENT
Option:
not applicable
Synchronization Required:
Preset State:
STARt
SCPI Compliance:
no
confirmed
[SENSe:]
command/query
18-45

Advertisement

Table of Contents
loading

Table of Contents