Agilent Technologies 35670A Manual page 179

Dynamic signal analyzer gpib programming
Hide thumbs Also See for 35670A:
Table of Contents

Advertisement

CALCulate[1|2|3|4]:MARKer:FUNCtion:RESult?
Returns the result of the calculation for the currently selected marker function.
Query Syntax:
Example Statements:
Return Format:
Attribute Summary:
Description:
This command returns the value of the marker function result calculation. To specify trace box A, send
CALC1. To specify trace box B, send CALC2; trace box C, send CALC3; and trace box D, send
CALC4. The trace specifier defaults to trace box A if the specifier is not used.
For example, send CALC3:MARK:FUNC GMAR then send CALC:MARK:FUNC:RES? to query the
value of the gain margin of trace C. To determine the units, send CALC3:MARK:FUNC:RES? UNIT.
Refer to the CALC:MARK:FUNCtion command for a complete listing of the available marker functions.
Caution
The analyzer returns the result 9.91E37, if it can not calculate a marker function.
In addition to function calculations, use this command to determine if one of the filters (A-weight,
B-weight, or C-weight) was applied to the measurement data or to determine the Z-axis value for the
measurement data.
CALCulate[1|2|3|4]:MARKer:FUNCtion:RESult?
OUTPUT 711;"CALCULATE4:MARK:FUNCTION:RES?"
OUTPUT 711;"calc4:marker:func:result?"
Real
Option:
not applicable
Synchronization Required:
Preset State:
not applicable
SCPI Compliance:
no
instrument-specific
CALCulate
query
6-71