CALCulate
CALCulate[1|2|3|4]:MARKer:BAND:STARt
Specifies the lowest frequency of the band in which power is calculated.
Command Syntax:
<number> ::= a real number (NRf data)
<unit> ::= [HZ|S|ORD|COUNT|AVG|V|VPK|RPM|EU|REV]
<step> ::= UP|DOWN
<bound> ::= MAX|MIN
Example Statements:
Query Syntax:
Return Format:
Attribute Summary:
Description:
This command defines the start value for the band used in calculating marker functions selected with the
CALC:MARK:FUNC command. The specified value affects only the currently selected marker function.
The value specified with the CALC:MARK:BAND:STAR command must be less than the value specified
with the CALC:MARK:BAND:STOP command.
Note
If you want to move the frequency band below the current frequency band, you must
reset the start value first. If you want to move the frequency band above the current
frequency band, you must reset the stop value first.
To increment the value to the next largest point on the X-axis, send CALC:MARK:BAND:STAR UP.
To decrement the value to the next smallest point on the X-axis, send CALC:MARK:BAND:STAR
DOWN.
6-52
CALCulate[1|2|3|4]:MARKer:BAND:STARt {<number>[<unit>]}|
<step>|<bound>
limits:
-9.9e37:9.9e37
OUTPUT 711;"calculate3:mark:band:star 0.0025 s"
OUTPUT 711;"CALC:MARK:BAND:START 10000 HZ"
CALCulate[1|2|3|4]:MARKer:BAND:STARt?
Real
Option:
not applicable
Synchronization Required:
Preset State:
not applicable
SCPI Compliance:
no
instrument-specific
command/query