CALCulate[1|2|3|4]:CFIT:COPY
Copies the synthesis table into the curve-fit table.
Command Syntax:
Example Statements:
Attribute Summary:
Description:
The synthesis table must be in pole-zero format. To convert a synthesis table to pole-zero format, use the
CALC:SYNT:TTYP PZER command.
Note
This command is not trace specific. It ignores the trace specifier.
CALCulate[1|2|3|4]:CFIT:COPY SYNThesis
OUTPUT 711;"calc:cfit:copy SYNTHESIS"
OUTPUT 711;"Calculate:Cfit:Copy SYNTHESIS"
Option:
1D3 Curve Fit/Synthesis
Synchronization Required:
Preset State:
not applicable
SCPI Compliance:
no
instrument-specific
CALCulate
command
6-5