TRACe
TRACe[:DATA]
Stores data to the specified data register.
Command Syntax:
<DATA> ::= D1|D2|D3|D4|D5|D6|D7|D8
<number> ::= <integer>
Example Statements:
See example program segments in Appendix F.
Query Syntax:
Return Format:
If FORMat[:DATA] REAL:
<DEF_BLOCK> ::= #<byte><length_bytes> <1st_value>[. . .<last_value>]
<byte> ::= one byte specifying the number of
<length_bytes> ::= number of data bytes to follow
If FORMat[:DATA] ASCii:
<DEF_BLOCK> ::= <1st_value>[. . .<last_value>]
Attribute Summary:
23-2
TRACe[:DATA] {D1|D2|D3|D4|D5|D6|D7|D8}, <DATA>
::= TRACe1|TRACe2|TRACe3|TRACe4
::= CAL1|CAL2|CAL3|CAL4
Calibration trace channel 1, 2, 3 or 4
::= <BLOCK> (definite and indefinite length)
::= <number>,<number>,<number> . . .
::= <floating-point numeric>
OUTPUT 711;"TRAC:DATA D1, TRAC2"
OUTPUT 711;"trac d3, d5"
TRACe[:DATA]? {D1|D2|D3|D4|D5|D6|D7|D8}
<DEF_BLOCK>
to follow (ASCII encoded)
Option:
not applicable
Synchronization Required:
Preset State:
not applicable
SCPI Compliance:
(ASCII encoded)
no
confirmed
command/query
length bytes