Calculate[1|2|3|4]:Synthesis:ttypecommand/Query - Agilent Technologies 35670A Manual

Dynamic signal analyzer gpib programming
Hide thumbs Also See for 35670A:
Table of Contents

Advertisement

CALCulate[1|2|3|4]:CFIT:ORDer:POLes
Specifies the number of poles used in the curve fit operation.
Command Syntax:
<number> ::= a real number (NRf data)
<step> ::= UP|DOWN
<bound> ::= MAX|MIN
Example Statements:
Query Syntax:
Return Format:
Attribute Summary:
Description:
The actual number of poles used in the curve fit operation is determined by the mode selected by the
CALC:CFIT:ORD:AUTO command.
If CALC:CFIT:ORD:AUTO is ON (the default value), an optimum number of poles is used for the
model. The number specified with this command represents the upper bounds for the iterative search.
The number of poles will not exceed the number specified by CALC:CFIT:ORD:POL.
If CALC:CFIT:ORD:AUTO OFF is sent, the number of poles specified by CALC:CFIT:ORD:POL is
used for the model.
Note
This command is not trace specific. It ignores the trace specifier.
CALCulate[1|2|3|4]:CFIT:ORDer:POLes <number>|<step>|
<bound>
limits:
0:20
OUTPUT 711;":calc:cfit:ord:poles 5"
OUTPUT 711;"Calc:Cfit:Order:Pol 14"
CALCulate[1|2|3|4]:CFIT:ORDer:POLes?
Integer
Option:
1D3 Curve Fit/Synthesis
Synchronization Required:
Preset State:
+20
SCPI Compliance:
no
instrument-specific
CALCulate
command/query
6-15

Advertisement

Table of Contents
loading