Calculate[1|2|3|4]:Unit:voltagecommand/Query - Agilent Technologies 35670A Manual

Dynamic signal analyzer gpib programming
Hide thumbs Also See for 35670A:
Table of Contents

Advertisement

CALCulate
To select the coherence function, send CALC[1|2|3|4]:FEED ' XFR:POW:COH [1,2|1,3|1,4|3,4]'.
The analyzer must be in 2 channel instrument mode (INPut 2 ON) for selection 1,2. INPut 4 must
be ON for selections 1,3 and 1,4. INPut 4 must be ON and SENse:REFerence must be PAIR for
selection 3,4.
To select the cross spectrum, send CALC[1|2|3|4]:FEED ' XFR:POW:CROS [1,2|1,3|1,4|3,4]'. The
analyzer must be in 2 channel instrument mode (INPut 2 ON) for selection 1,2. INPut 4 must be
ON for selections 1,3 and 1,4. INPut 4 must be ON and SENse:REFerence must be PAIR for
selection 3,4.
To select the most recent frequency response function, send CALC[1|2|3|4]:FEED 'XFR:POW:RAT
[2,1|3,1|4,1|4,3]'. The analyzer must be in 2 channel instrument mode (INPut 2 ON) for selection
2,1. INPut 4 must be ON for selections 3,1 and 4,1. INPut 4 must be ON and SENse:REFerence
must be PAIR for selection 4,3.
To select the most recent time record for the specified channel, send CALC[1|2|3|4]:FEED
'XTIM:VOLT [1|2|3|4]'.
To select the most recent windowed time record for the specified channel, send
CALC[1|2|3|4]:FEED 'XTIM:VOLT:WIND [1|2|3|4]'.
To select the orbit diagram, send CALC[1|2|3|4]:FEED 'XVOL:VOLT [1,2|1,3|1,4|3,4]'. The orbit
diagram presents channel 1 time along the X-axis and channel 2, channel 3, or channel 4 time along
the Y- axis. The analyzer must be in 2 channel instrument instrument mode (INPut 2 ON) for
selection 1,2. INPut 4 must be ON for selections 1,3 and 1,4. INPut 4 must be ON and
SENse:REFerence must be PAIR for selection 3,4.
In correlation instrument mode (INST:SEL CORR), the following commands are available:
CALC[1|2|3|4]:FEED 'D1|D2|D3|D4|D5|D6|D7|D8'.
CALC[1|2|3|4]:FEED 'TCAP [1|2|3|4]'.
CALC[1|2|3|4]:FEED 'W1|W2|W3|W4|W5|W6|W7|W8'.
CALC[1|2|3|4]:FEED 'XTIM:VOLT [1|2|3|4]'.
CALC[1|2|3|4]:FEED 'XTIM:VOLT:WIND [1|2|3|4]'.
To select the most recent autocorrelation for the specified channel, send CALC[1|2|3|4]:FEED
'XTIM:CORR [1|2|3|4]'.
To select the most recent cross correlation for the specified channel, send CALC[1|2|3|4]:FEED
'XTIM:CORR:CROS [1,2|1,3|1,4|3,4]'. The analyzer must be in 2 channel instrument mode (INPut
2 ON) for selection 1,2. INPut 4 must be ON for selections 1,3 and 1,4. INPut 4 must be ON and
SENse:REFerence must be PAIR for selection 3,4.