Agilent Technologies 35670A Manual page 547

Dynamic signal analyzer gpib programming
Hide thumbs Also See for 35670A:
Table of Contents

Advertisement

SOURce:VOLTage:LIMit[:AMPLitude]
Sets the maximum limit used by the autolevel algorithm to adjust the source's amplitude in swept sine
measurements (INST:SEL SINE).
Command Syntax:
<number> ::= a real number (NRf data)
<unit> ::= [DBVRMS|VPK|DBVPK|V|DBV|VRMS]
<step> ::= UP|DOWN
<bound> ::= MAX|MIN
Example Statements:
Query Syntax:
Return Format:
Attribute Summary:
Description:
This command controls the analyzer's autolevel algorithm in adjusting the source's output level during a
sweep. The limit constrains any autolevel adjustment to the source's amplitude.
SOURce:VOLTage:LIMit[:AMPLitude] {<number>
[<unit>]}|<step>|<bound>
limits:
-9.9e37:13.9794
OUTPUT 711;"Source:Voltage:Limit 0 DBV"
OUTPUT 711;"sour:volt:lim 1 VRMS"
SOURce:VOLTage:LIMit[:AMPLitude]?
Real
Option:
1D2 Swept Sine
Synchronization Required:
Preset State:
+2.0 VPK
SCPI Compliance:
no
confirmed
SOURce
command/query
19-15

Advertisement

Table of Contents
loading

Table of Contents