[SENSe:]
In histogram analysis instrument mode (INST:SEL HIST):
This command specifies the length of time averaging data for the histogram. This histogram length can be
specified in time (S for seconds), records (REC), or points (PNT). The analyzer rounds the specified
histogram length up to the nearest point.
An optimal histogram may be obtained by setting the number of points (specified by this command) to the
number of bins2 (specified with the HIST:BINS command).
\Optimal Histogram
AVER TIME
=
HIST BINS
18-18