Agilent Technologies 35670A Manual page 195

Dynamic signal analyzer gpib programming
Hide thumbs Also See for 35670A:
Table of Contents

Advertisement

CALCulate[1|2|3|4]:MARKer:SIDeband:COUNt
Specifies the number of sideband markers for the display.
Command Syntax:
<number> ::= a real number (NRf data)
<step> ::= UP|DOWN
<bound> ::= MAX|MIN
Example Statements:
Query Syntax:
Return Format:
Attribute Summary:
Description:
The value can be specified numerically or with nonnumeric parameters.
To increase the number of sideband markers by one, send CALC:MARK:HARM:COUN UP.
To decrease the number of sideband markers by one, send CALC:MARK:HARM:COUN DOWN.
The query (CALC:MARK:HARM:COUNT?) returns the number of sideband markers currently specified
for the display. The value is returned even if the sideband markers are not on.
CALCulate[1|2|3|4]:MARKer:SIDeband:COUNt <number>|<step>|
<bound>
limits:
0:200
OUTPUT 711;":calc4:marker:sid:count 2"
OUTPUT 711;"CALC:MARK:SID:COUN 6"
CALCulate[1|2|3|4]:MARKer:SIDeband:COUNt?
Integer
Option:
not applicable
Synchronization Required:
Preset State:
not applicable
SCPI Compliance:
no
instrument-specific
CALCulate
command/query
6-87