HCOPy:PLOT:ADDRess
Tells the analyzer which GPIB address is assigned to your plotter.
Command Syntax:
<number> ::= a real number (NRf data)
<step> ::= UP|DOWN
<bound> ::= MAX|MIN
Example Statements:
Query Syntax:
Return Format:
Attribute Summary:
Description:
This command works when the HCOP:DESTination command is set to GPIB and the HCOP:DEVice
command is set to PLOT.
Initiate a plot with the HCOPy[:IMM] command. The analyzer expects to find a plotter at the GPIB
address specified with HCOP:PLOT:ADDR. If a plotter is not at the specified address, the plot is
automatically aborted.
HCOPy:PLOT:ADDRess <number>|<step>|<bound>
limits:
0:30
OUTPUT 711;"HCOPY:PLOT:ADDRESS 5"
OUTPUT 711;"hcop:plot:addr 9"
HCOPy:PLOT:ADDRess?
Integer
Option:
not applicable
Synchronization Required:
Preset State:
not affected by Preset
SCPI Compliance:
no
instrument-specific
HCOPy
command/query
10-23