[SENSe:]TCAPture[:IMMediate]
Starts the collection of data for the time capture process.
Command Syntax:
Example Statements:
Attribute Summary:
Description:
The analyzer collects data from the input channels and stores it in the time capture buffer.
If the command (TCAP:MALL) which allocates memory has not been sent; the analyzer automatically
allocates memory based on the current settings for the frequency span, the size of the time capture buffer
(TCAP:LENG) and the size of the tachometer buffer if it is enabled with the
[SENSe:]TCAP:TACH[:STATe] ON command.
The size of the tachometer buffer is determined by the number of revolutions per minute specified with
the [SENSe:]TCAP:TACH:RPM:MAX command and the number of tachometer pulses per revolution
specified with the TRIGger:TACH:PCOunt command.
[SENSe:]TCAPture[:IMMediate]
OUTPUT 711;"SENS:TCAP:IMMEDIATE"
OUTPUT 711;"tcap"
Option:
not applicable
Synchronization Required:
Preset State:
not applicable
SCPI Compliance:
no
instrument-specific
[SENSe:]
command
18-71