Agilent Technologies 35670A Manual page 219

Dynamic signal analyzer gpib programming
Hide thumbs Also See for 35670A:
Table of Contents

Advertisement

CALCulate[1|2|3|4]:UNIT:AMPLitude
Selects the unit of amplitude for the Y-axis scale.
Command Syntax:
Example Statements:
Query Syntax:
Return Format:
Attribute Summary:
Description:
To display peak amplitude, send CALC:UNIT:AMPL PEAK.
To display peak-to-peak amplitude, send CALC:UNIT:AMPL PP.
To display RMS amplitude, send CALC:UNIT:AMPL RMS.
The default value is dependent upon the selected measurement data (CALC:FEED). Table 6-6 indicates
valid unit selections for the CALC:UNIT:AMPL command. If measurement data does not appear in the
table, you are not permitted to select the amplitude. In this case, a query returns a null string. See
"Determining Units" in Appendix E for information about available Y-axis units.
CALCulate[1|2|3|4]:UNIT:AMPLitude PEAK|PP|RMS
OUTPUT 711;"CALC3:UNIT:AMPL PP"
OUTPUT 711;"calculate4:unit:ampl PP"
CALCulate[1|2|3|4]:UNIT:AMPLitude?
PEAK|PP|RMS
Option:
not applicable
Synchronization Required:
Preset State:
RMS (Channel 1 and Channel 3)
PEAK (Channel 2 and Channel 4)
SCPI Compliance:
yes
instrument-specific
CALCulate
command/query
6-111