CALCulate
CALCulate[1|2|3|4]:LIMit:LOWer:MOVE:Y
Moves all segments of the lower limit line up or down in the specified trace.
Command Syntax:
<number> ::= a real number (NRf data)
<bound> ::= MAX|MIN
Example Statements:
Attribute Summary:
Description:
This command increments or decrements all segments in a lower limit by the specified value along the
Y-axis. The value is unitless and assumes the current vertical/division unit (returned with
DISP:TRAC:Y:PDIV? UNIT).
To specify trace box A, send CALC1:LIM:LOW:MOVE:Y. To specify trace box B, send
CALC2:LIM:LOW:MOVE:Y. To specify trace box C, send CALC3:LIM:LOW:MOVE:Y. To specify
trace box D, send CALC4:LIM:LOW:MOVE:Y. If a trace specifier is not included in the command, the
trace specifier defaults to trace box A.
6-36
CALCulate[1|2|3|4]:LIMit:LOWer:MOVE:Y <number>|<bound>
limits: -9.9e37:9.9e37
(depend upon current verticall/division unit)
OUTPUT 711;"Calculate4:Lim:Lower:Move:Y 9.55849e+37"
OUTPUT 711;"CALCULATE:LIM:LOWER:MOVE:Y -3.82565e+37"
Option:
not applicable
Synchronization Required:
Preset State:
not applicable
SCPI Compliance:
no
instrument-specific
command