Calculate[1|2|3|4]:Marker:maximum[:Global]Command - Agilent Technologies 35670A Manual

Dynamic signal analyzer gpib programming
Hide thumbs Also See for 35670A:
Table of Contents

Advertisement

Common Commands
*STB?
Reads the Status Byte register.
Query Syntax:
Example Statements:
Return Format:
Attribute Summary:
Description:
This command allows you to set bits in the Status Byte register. The state is returned as a sum of the
decimal weights of all set bits. The decimal weight for each bit is assigned according to the following
formula:
with acceptable values for bit_number being 0 through 7.
The register is not cleared by this query. To clear the Status Byte register, you must send the *CLS
command.
Bits in the Status Byte register are defined as follows:
Bit 0 summarizes all enabled bits of the User Status register.
Bit 1 is reserved.
Bit 2 summarizes all enabled bits of the Device State register.
Bit 3 summarizes all enabled bits of the Questionable Status register.
Bit 4 is the Message Available (MAV) bit. It is set whenever there is something in the analyzer's
output queue.
Bit 5 summarizes all enabled bits of the Standard Event Status register.
3-16
*STB?
OUTPUT 711;"*Stb?"
OUTPUT 711;"*STB?"
Integer
Option:
not applicable
Synchronization Required:
Preset State:
variable
SCPI Compliance:
(bit_number)
2
no
confirmed
query