[SENSe:]
[SENSe:]FEED
Specifies the data source for a measurement; either from the input channels or from the time capture
buffer.
Command Syntax:
Example Statements:
Query Syntax:
Return Format:
Attribute Summary:
Description:
Data from the time capture buffer may be used for a measurement, replacing the use of the inputs
(channels 1, 2, 3, and 4). This command directs the analyzer to use the data in the time capture buffer for
the measurement.
Note
The measurement does not begin until the INITiate command is sent.
18-26
[SENSe:]FEED INPut|TCAPture
OUTPUT 711;":Feed TCAPTURE"
OUTPUT 711;"SENS:FEED TCAPTURE"
[SENSe:]FEED?
INP|TCAP
Option:
not applicable
Synchronization Required:
Preset State:
INP
SCPI Compliance: