Agilent Technologies 35670A Manual page 580

Dynamic signal analyzer gpib programming
Hide thumbs Also See for 35670A:
Table of Contents

Advertisement

STATus
STATus:USER:ENABle
Sets and queries bits in the User Defined enable register.
Command Syntax:
<number> ::= a real number (NRf data)
<bound> ::= MAX|MIN
Example Statements:
Query Syntax:
Return Format:
Attribute Summary:
Description:
To set a single bit in the User Defined enable register to 1, send the bit's decimal weight with this
command. To set more than one bit to 1, send the sum of the decimal weights of all the bits. (The
decimal weight of a bit is 2
All bits are initialized to 0 when the analyzer is turned on or when the STAT:PRES command is sent.
However, the current setting of bits is not modified when you send the *RST command.
See "User Defined Register Set" in chapter 1 for a definition of bits in the register set. See
"Programming the Status System" in the GPIB Programmer's Guide for information about the role
of enable registers in register sets.
20-28
STATus:USER:ENABle <number>|<bound>
limits:
0:32767
OUTPUT 711;"Status:User:Enab 18066"
OUTPUT 711;"STATUS:USER:ENABLE 10359"
STATus:USER:ENABle?
Integer
Option:
not applicable
Synchronization Required:
Preset State:
not affected by Preset
SCPI Compliance:
n
, where n is the bit number.)
no
instrument-specific
command/query

Advertisement