Agilent Technologies 35670A Manual page 336

Dynamic signal analyzer gpib programming
Hide thumbs Also See for 35670A:
Table of Contents

Advertisement

HCOPy
HCOPy:PAGE:DIMensions:USER:URIGht
Specifies the upper right position (P2) of the plot area.
Command Syntax:
<P2_X-axis_value> ::= <number>|<step>|<bound>
<number> ::= a real number (NRf data)
<step> ::= UP|DOWN
<bound> ::= MAX|MIN
<P2_Y-axis_value> ::= <number>|<step>|<bound>
<number> ::= a real number (NRf data)
<step> ::= UP|DOWN
<bound> ::= MAX|MIN
Example Statements:
Query Syntax:
Return Format:
Attribute Summary:
Description:
The plot area is defined by the X-axis and Y-axis values for the scaling points P1 and P2. The X-axis
value and the Y-axis value depends upon the plotter and the number of points on the page.
Use the HCOP:PAGE:DIM:USER:LLEF command to specify the P1 X- and Y-axis values.
Check the documentation for your plotter to select appropriate P2 values.
Note
HCOP:PAGE:DIM:USER:URIG is set to +9155,+7377 after a reset (*RST).
10-22
HCOPy:PAGE:DIMensions:USER:URIGht {<P2_X-axis_value>,
<P2_Y-axis_value>}
limits:
-32767:32767
limits:
-32767:32767
OUTPUT 711;"HCOP:PAGE:DIM:USER:URIG 4743,7377"
OUTPUT 711;"hcopy:page:dimensions:user:uright 9155,4286"
HCOPy:PAGE:DIMensions:USER:URIGht?
Integer, Integer
Option:
not applicable
Synchronization Required:
Preset State:
not affected by Preset
SCPI Compliance:
no
instrument-specific
command/query

Advertisement

Table of Contents
loading