Agilent Technologies 35670A Manual page 510

Dynamic signal analyzer gpib programming
Hide thumbs Also See for 35670A:
Table of Contents

Advertisement

[SENSe:]
[SENSe:]TCAPture:LENGth
Specifies the length of the time capture buffer.
Command Syntax:
<number> ::= a real number (NRf data)
<unit> ::= [S|BLK|PNT]
<step> ::= UP|DOWN
<bound> ::= MAX|MIN
Example Statements:
Query Syntax:
Return Format:
Attribute Summary:
Description:
If length is specified in seconds, the size of the buffer is relative to the frequency span.
If length is specified in blocks or points, the size of the buffer is absolute. There are 1024 sample points
per block.
Any specified length is rounded up to the nearest block.
The maximum capture length is dependent upon the memory configuration:
18-72
[SENSe:]TCAPture:LENGth {<number>[<unit>]}|<step>|<bound>
limits:
0:9.9e37
OUTPUT 711;":Tcapture:Leng 4.54329e+36"
OUTPUT 711;"SENS:TCAPTURE:LENG 2.91628e+37"
[SENSe:]TCAPture:LENGth?
Real
Option:
not applicable
Synchronization Required:
Preset State:
+0.0
SCPI Compliance:
no
instrument-specific
command/query

Advertisement

Table of Contents
loading