[SENSe:]ORDer:RPM:MINimum
Specifies the minimum rotational speed range you want to analyze.
Command Syntax:
<number> ::= a real number (NRf data)
<unit> ::= [RPM]
<step> ::= UP|DOWN
<bound> ::= MAX|MIN
Example Statements:
Query Syntax:
Return Format:
Attribute Summary:
Description:
For runup measurements, the measurement starts at the speed specified by this command and continues to
the maximum RPM (ORD:RPM:MAX).
For rundown measurements, the measurement stops at the speed specified by this command.
If the value you specify for ORD:RPM:MIN is greater than the value set for ORD:RPM:MAX, the value
for ORD:RPM:MAX is set to equal the value of ORD:RPM:MIN.
If
ORD:RPM:MIN > ORD:RPM:MAX
then
ORD:RPM:MIN = ORD:RPM:MAX
If the value you specify for ORD:RPM:MAX is less than the value set for ORD:RPM:MIN, the value for
ORD:RPM:MIN is set to equal the value of ORD:RPM:MAX.
If
ORD:RPM:MAX < ORD:RPM:MIN
then
ORD:RPM:MIN = ORD:RPM:MAX
[SENSe:]ORDer:RPM:MINimum {<number>[<unit>]}|<step>|
<bound>
limits:
0:9.9e37
OUTPUT 711;"SENSE:ORDER:RPM:MINIMUM 800"
OUTPUT 711;"ord:rpm:min 1E4"
[SENSe:]ORDer:RPM:MINimum?
Real
Option:
1D0
Computed Order Tracking
Synchronization Required:
Preset State:
+600
SCPI Compliance:
no
instrument-specific
[SENSe:]
command/query
18-55