Calculate[1|2|3|4]:Synthesis[:Immediate]Command - Agilent Technologies 35670A Manual

Dynamic signal analyzer gpib programming
Hide thumbs Also See for 35670A:
Table of Contents

Advertisement

CALCulate
CALCulate[1|2|3|4]:CFIT:FSCale
Specifies the frequency scaling used in the curve fit operation.
Command Syntax:
<number> ::= a real number (NRf data)
<step> ::= UP|DOWN
<bound> ::= MAX|MIN
Example Statements:
Query Syntax:
Return Format:
Attribute Summary:
Description:
The analyzer scales the frequency axis (the X-axis) by f/frequency scale, where f is frequency in Hertz.
Note
This command must be sent before CALC:CFIT. This command is not trace specific. It
ignores the trace specifier.
6-12
CALCulate[1|2|3|4]:CFIT:FSCale <number>|<step>|<bound>
limits:
1e-6:1e6
OUTPUT 711;"calculate2:cfit:fscale 604144"
OUTPUT 711;"Calc2:Cfit:Fscale 582699"
CALCulate[1|2|3|4]:CFIT:FSCale?
Real
Option:
1D3 Curve Fit/Synthesis
Synchronization Required:
Preset State:
+1.0
SCPI Compliance:
no
instrument-specific
command/query