DISPlay[:WINDow[1|2|3|4]]:WATerfall:SKEW:ANGLe
Specifies the amount of horizontal offset for waterfall displays.
Command Syntax:
<number> ::= a real number (NRf data)
<step> ::= UP|DOWN
<bound> ::= MAX|MIN
Example Statements:
Query Syntax:
Return Format:
Attribute Summary:
Description:
This command is used with the DISPlay:WATerfall:SKEW command. In a waterfall display, a trace is
offset from the horizontal axis by the "angle of skew" specified with this command. The trace scrolls
down the display towards the origin. See figure 8-3.
WINDow[1|2|3|4] specifies the trace box. Send WIND1 to specify A, WIND2 to specify B, WIND3 to
specify C, and WIND4 to specify D. If you do not send the optional trace box specifier, the command
defaults to trace box A.
DISPlay[:WINDow[1|2|3|4]]:WATerfall:SKEW:ANGLe <num-
ber>|<step>|<bound>
limits:
-45:45
OUTPUT 711;"disp:wat:skew:angl 30"
OUTPUT 711;"DISP:WIND3:WAT:SKEW:ANGLE 20"
DISPlay[:WINDow[1|2|3|4]]:WATerfall:SKEW:ANGLe?
Integer
Option:
not applicable
Synchronization Required:
Preset State:
30°
SCPI Compliance:
Figure 8-3. Waterfall Traces Displayed
at 45⊃ Angle of Skew
no
instrument-specific
DISPlay
command/query
8-49