Agilent Technologies 35670A Manual page 542

Dynamic signal analyzer gpib programming
Hide thumbs Also See for 35670A:
Table of Contents

Advertisement

SOURce
SOURce:VOLTage[:LEVel][:IMMediate][:AMPLitude]
Specifies the source output level.
Command Syntax:
<number> ::= a real number (NRf data)
<unit> ::= [DBVRMS|VPK|DBVPK|V|DBV|VRMS]
<step> ::= UP|DOWN
<bound> ::= MAX|MIN
Example Statements:
Query Syntax:
Return Format:
Attribute Summary:
Description:
The source output level can be expressed in terms of peak values, Vpk, or rms values, Vrms.
While most source waveforms are limited to a maximum of 5 VPK, pink noise (SOUR:FUNC PINK) is
limited to 4.196 VPK. The minimum is 0 VPK. If the source output level is specified in DBVRMS, the
smallest non-zero level is -74.912 DBVRMS.
To determine if the output level is set to PEAK, PP (peak-to-peak), or RMS values, send SOUR:VOLT?
UNIT. Each source waveform has a specific peak-to-rms value. The maximum rms level depends upon
the source waveform. See Online Help for additional information.
19-10
SOURce:VOLTage[:LEVel][:IMMediate][:AMPLitude] {<number>
[<unit>]}|<step>|<bound>
limits:
-9.9e37:13.9794
OUTPUT 711;"SOUR:VOLT 0 DBV"
OUTPUT 711;"sour:volt:lev:imm:amp 5 VPK"
SOURce:VOLTage[:LEVel][:IMMediate][:AMPLitude]?
Real
Option:
not applicable
Synchronization Required:
Preset State:
+0.0 VPK
SCPI Compliance:
no
confirmed
command/query

Advertisement

Table of Contents
loading

Table of Contents