SYSTem:BEEPer:STATe
Enables the analyzer's beeper.
Command Syntax:
Example Statements:
Query Syntax:
Return Format:
Attribute Summary:
Description:
When the beeper is enabled, it emits an audible tone when some messages are either displayed or placed
in the error queue. It also emits an audible tone when a trace falls outside its specified limits if limit
testing and the limit-fail beeper is turned on (CALCulate:LIMit:STATe ON and CALCulate:LIMit:BEEP
ON).
When the beeper is disabled, the SYST:BEEP command will not generate any sound.
SYSTem:BEEPer:STATe OFF|0|ON|1
OUTPUT 711;":syst:beep:state ON"
OUTPUT 711;"Syst:Beeper:Stat OFF"
SYSTem:BEEPer:STATe?
Integer
Option:
not applicable
Synchronization Required:
Preset State:
ON (+1)
SCPI Compliance:
no
confirmed
SYSTem
command/query
21-3