Calculate[1|2|3|4]:Marker:sideband:incrementcommand/Query - Agilent Technologies 35670A Manual

Dynamic signal analyzer gpib programming
Hide thumbs Also See for 35670A:
Table of Contents

Advertisement

ARM:RPM:INCRement
Specifies the number of RPM in a step for RPM step arming.
Command Syntax:
<number> ::= a real number (NRf data)
<unit> ::= [RPM]
<step> ::= UP|DOWN
<bound> ::= MAX|MIN
Example Statements:
Query Syntax:
Return Format:
Attribute Summary:
Description:
This command determines the size of each step used when RPM step arming is enabled with the
ARM:SOUR RPM command.
The first arm occurs when the RPM value reaches the threshold value (specified with the
ARM:RPM:THR command). Subsequent arms occur at the RPM interval specified with this command.
ARM:RPM:INCRement {<number>[<unit>]}|<step>|<bound>
limits:
1:500000
OUTPUT 711;"ARM:RMP:INCR 10"
OUTPUT 711;"arm:rmp:increment 500"
ARM:RPM:INCRement?
Real
Option:
not applicable
Synchronization Required:
Preset State:
+60 RPM
SCPI Compliance:
no
instrument-specific
ARM
command/query
5-3

Advertisement

Table of Contents
loading