DISPlay[:WINDow[1|2|3|4]]:WATerfall:COUNt
Determines the number of traces displayed in the waterfall trace box.
Command Syntax:
<number> ::= a real number (NRf data)
<unit> ::= [S|RPM|COUNT|AVG]
<step> ::= UP|DOWN
<bound> ::= MAX|MIN
Example Statements:
Query Syntax:
Return Format:
Attribute Summary:
Description:
The number of traces displayed in a waterfall is determined by the range of Z-axis values specified with
this command. The analyzer may adjust the specified range to include the trace selected with the
CALC:WAT:TRAC:SEL command.
WINDow[1|2|3|4] specifies the trace box. Send WIND1 to specify A, WIND2 to specify B, WIND3 to
specify C, and WIND4 to specify D. If you do not send the optional trace box specifier, the command
defaults to trace box A.
Note
All traces in a waterfall display are deleted when the ABORT;:INIT:CONT ON
command is sent.
DISPlay[:WINDow[1|2|3|4]]:WATerfall:COUNt {<num-
ber>[<unit>]}|<step>|<bound>
limits:
1e-6:9.9e37
OUTPUT 711;":Display:Wat:Coun 5"
OUTPUT 711;"DISP:WIND2:WAT:COUNT 100 RPM"
DISPlay[:WINDow[1|2|3|4]]:WATerfall:COUNt?
Real
Option:
not applicable
Synchronization Required:
Preset State:
+14 COUNT
SCPI Compliance:
no
instrument-specific
DISPlay
command/query
8-45

Advertisement