MEMory:CATalog:NAME?
Returns information about memory usage allocated for a specific item.
Query Syntax:
Example Statements:
Return Format:
<NAME> ::= TCAP (time capture buffer)
Attribute Summary:
Description:
This query returns the amount of the memory allocated for a specific item.
To allocate memory for a time capture buffer, send the [SENSe:]TCAPture:MALLocate command.
To allocate memory for a waterfall display, send the CALCulate:WATerfall:COUNt command.
The analyzer allocates memory for a waterfall register when you send the TRACe:WATerfall[:DATA]
command.
The analyzer allocates memory for an Instrument BASIC program when you create the program. The
memory usage is allocated for all Instrument BASIC programs. If multiple programs reside within the
analyzer, you can not determine the memory allocated for a single Instrument BASIC program with this
command.
To allocate memory for the volatile RAM disk, send the MMEMory:INITialize command.
The query returns <NAME> in short form (for example, TCAP for time capture).
MEMory:CATalog:NAME? TCAPture|WATerfall|WREGister|PRO-
Gram|RDISk|LIMits
OUTPUT 711;"MEMORY:CAT:NAME? WREGISTER"
OUTPUT 711;"mem:cat:name? wat"
<NAME>
WAT (waterfall)
WREG (waterfall register)
PROG (Instrument BASIC program)
RDIS (volatile RAM disk)
Option:
not applicable
Synchronization Required:
Preset State:
not applicable
SCPI Compliance:
no
instrument-specific
MEMory
query
14-3