[SENSe:]
[SENSe:]VOLTage[1|2|3|4][:DC]:RANGe:UNIT:USER[:STATe]
Enables the use of transducer units.
Command Syntax:
Example Statements:
Query Syntax:
Return Format:
Attribute Summary:
Description:
This command allows you to specify a transducer unit for each input channel.
When VOLT:RANG:UNIT:USER is OFF, the unit is Volts.
When VOLT:RANG:UNIT:USER is ON, the transducer unit is specified by the
VOLT:RANG:UNIT:XDCR command and the transducer sensitivity is specified by the
VOLT:RANG:UNIT:USER:SFACtor command. If the channel specifier is not used, the command
defaults to channel 1.
To use transducer units:
1.Specify a unit label.
a. To select a label for sound pressure, acceleration, velocity, or displacement, use the
VOLT:RANG:UNIT:XDCR:LABel command. Parameters include Pascals, g's, meters, meters
per second, meters per second
per second
2
, pounds, and mils.
b. To define your own label, set VOLT:RANG:UNIT:XDCR:LABel to USER and send the
VOLT:RANG:UNIT:LABel command with your own label.
2.Specify the transducer's sensitivity with the VOLT:RANG:UNIT:USER:SFACtor command.
3.Enable the transducer unit setup with VOLT:RANG:UNIT:USER ON.
18-84
[SENSe:]VOLTage[1|2|3|4][:DC]:RANGe:UNIT:USER[:STATe]
OFF|0|ON|1
OUTPUT 711;":voltage:rang:unit:user OFF"
OUTPUT 711;"Sens:Voltage:Dc:Rang:Unit:User:State OFF"
[SENSe:]VOLTage[1|2|3|4][:DC]:RANGe:UNIT:USER[:STATe]?
Integer
Option:
not applicable
Synchronization Required:
Preset State:
OFF (+0)
SCPI Compliance:
2
, kilograms, Newtons, dynes, inches, inches per second, inches
no
instrument-specific
command/query