[SENSe:]AVERage:COUNt
Specifies a count or a weighting factor for the averaged measurement data.
Command Syntax:
<number> ::= a real number (NRf data)
<step> ::= UP|DOWN
<bound> ::= MAX|MIN
Example Statements:
Query Syntax:
Return Format:
Attribute Summary:
Description:
As a counter, AVER:COUN determines the number of time records used to average the measurement
data. This command is only valid in FFT, order analysis and correlation instrument modes.
Once the specified number of time records have been averaged, the No Pending Operation (NPO) flag is
set to 1. Use the *OPC command to determine when the specified number of time records have been
combined. (See "Synchronization" in the GPIB Programmer's Guide for more information about the
completion of averaged measurements.)
When used with exponential averaging (AVER:TYPE RMS;TCON EXP or AVER:TYPE TIME;TCON
EXP), this command determines how the results of the current measurement (new data) is combined with
the averaged trace (old data). Data is combined, point by point, according to the following formula:
[SENSe:]AVERage:COUNt <number>|<step>|<bound>
limits:
1:9999999
OUTPUT 711;"sense:average:count 2"
OUTPUT 711;"AVER:COUN 100"
[SENSe:]AVERage:COUNt?
Integer
Option:
not applicable
Synchronization Required:
Preset State:
+10
SCPI Compliance:
1
new
n
no
confirmed
n
1
+
old
n
[SENSe:]
command/query
18-3