[SENSe:]
[SENSe:]SWEep:STIMe
Specifies the settling time for a swept sine measurement.
Command Syntax:
<number> ::= a real number (NRf data)
<unit> ::= [S|CYCLE]
<step> ::= UP|DOWN
<bound> ::= MAX|MIN
Example Statements:
Query Syntax:
Return Format:
Attribute Summary:
Description:
Settling time is the delay between changing the source frequency and starting the measurement at each
point. This allows the device under test to stabilize after the frequency changes.
Units can be specified in seconds (S) or cycles (CYCLE).
18-66
[SENSe:]SWEep:STIMe {<number>[<unit>]}|<step>|<bound>
limits:
0:9.9e37
OUTPUT 711;"SENS:SWEEP:STIM 0.005 s"
OUTPUT 711;"swe:stime 1 cycle"
[SENSe:]SWEep:STIMe?
Real
Option:
1D2 Swept Sine
Synchronization Required:
Preset State:
+5.0 CYCLE
SCPI Compliance:
no
instrument-specific