INPut[1|2|3|4]:FILTer[:LPASs][:STATe]
Enables/disables the anti-alias filter for the specified input channel.
Command Syntax:
Example Statements:
Query Syntax:
Return Format:
Attribute Summary:
Description:
When INPut[1|2|3|4]:FILTer is OFF, the analyzer's corresponding input bypasses the anti-alias low pass
filter. Measurement results are not corrected for front end flatness; only front end DC offset is calibrated.
If the channel specifier is not used, the command defaults to channel 1.
With the Agilent 35670A Option AY6 (4 input channels), disabling the anti-alias filter in one channel
instrument mode (INP2 OFF), allows you to work with a 102.4 kHz frequency span.
Note
The anti-alias low pass filter is always bypassed in histogram instrument mode.
INPut[1|2|3|4]:FILTer[:LPASs][:STATe] OFF|0|ON|1
OUTPUT 711;":INP2:FILTER OFF"
OUTPUT 711;"inp:filt:lpass:stat ON"
INPut[1|2|3|4]:FILTer[:LPASs][:STATe]?
Integer
Option:
not applicable
Synchronization Required:
Preset State:
ON (+1)
SCPI Compliance:
no
confirmed