FREQ:SPAN UP increases the frequency span to the next largest allowable value. FREQ:SPAN
DOWN decreases the frequency span to the next smallest allowable value.
The frequency resolution is determined by the [SENSe:]FREQuency:RESolution command. The
frequency span is proportional to the sampling rate. To increase the sampling rate, you must
increase the frequency span.
In octave analysis instrument mode (INST:SEL OCT; Option 1D1):
This command is not valid. The frequency resolution is determined by the
[SENSe:]FREQuency:RESolution:OCTave command.
In order analysis instrument mode (INST:SEL ORD;Option 1D0):
This command is not valid. The frequency resolution is determined by the frequency span.
In swept sine instrument mode (INST:SEL SINE; Option 1D2):
The value specified by this command is rounded to the next higher 15.625 mHz step. The
frequency span limits the range of the start frequency (FREQ:STAR) values according to the
following formulas:
for two channel measurements:
(FREQ:STAR)  51.2 kHz - (FREQ:SPAN)
for four channel measurements:
(FREQ:STAR)  25.6 kHz - (FREQ:SPAN)
If logarithmic spacing is used (SWE:SPAC LOG), the value may be expressed in terms of hertz,
decades or octaves.
In correlation analysis instrument mode (INST:SEL CORR):
This command is not valid.